The biochemical incubator is a kind of teaching equipment, which is suitable for bioengineering, medical research, health and epidemic prevention, environmental protection, drug testing, animal husbandry, aquatic products, colleges and universities, research institutions and other fields. The biochemical incubator has the functions of cooling and heating control and is an indispensable device for the cultivation of microorganisms and bacteria.
Features:
1 box made of high quality, surface spray, the use of medical mirror stainless steel liner, easy to clean, the box shelf spacing adjustable, large viewing window;
2 microcomputer intelligent digital display temperature controller, PID control, touch type button, while displaying the set temperature and temperature inside the box, temperature control accuracy and reliability, with a timing function, and with over-temperature protection device, sound and light alarm function;
3 The use of brand compressors, environmentally friendly refrigerants, high efficiency, low energy consumption, and promote energy conservation.
4 brand micro-fans form a reasonable and advanced temperature cycle system, so that the box within a small test evenly.
There are 5 lights in the box to observe the culture in the box;
6 can be equipped with RS-485 interface, can be connected to the printer and computer, record changes in temperature parameters (optional)
Biochemical incubator instructions for use:
1. Turn on the power, close the power switch, and turn on the power. The power indicator in the switch is on.
2. Please refer to the manual of intelligent digital temperature controller for temperature control setting.
3, if you need to turn on the lighting switch.
4. The incubator has a break point protection function and a delay time of about one and a half minutes. Once the compressor stops, it will take about one and a half minutes to start again.
Biochemical incubator maintenance and culture:
1, incubator shell should be reliably grounded.
2. Incubator should be placed in a cool, dry, well ventilated place away from heat and sunlight. Keep it stable to prevent vibrations from generating noise.
3, in order to ensure that the condenser effectively cooling, the distance between the condenser and the wall should be greater than 100mm. There should be 50mm clearance on the side of the box, and there should be at least 300mm on the top of the box.
4. The incubator should avoid collisions and shake vibrations during handling, maintenance and maintenance; the maximum inclination is less than 45 degrees.
5, the instrument suddenly does not work, please check the fuse tube (after the box) is burned, check the power supply.
6. When the biochemical incubator is working in refrigeration, it should not make the difference between the temperature in the box and the ambient temperature more than 25 degrees.