Feature: This system is made up of seed fermentor and main fermentor, the two fermentors are controlled by one control system in on control cabinet. The whole system is installed on one platform, the two fermentors can work independently and can work together at same time.
It adopts top mechanical agitation system, the rotate speed can be adjustable. Parameters of temp. pH, DO, feed, foam, etc can be automatically controlled, recorded, stored and printed.
It has the following features: Excellent stability and convenient operation. Application: This series of products suit for various kinds of microbiology lab, pilot and industrial scale fermentation. It is ideal equipment for two step pilot scale fermentation. User can choose appreciate model according to client detail requirement.
It adopts top mechanical agitation system, the rotate speed can be adjustable. Parameters of temp. pH, DO, feed, foam, etc can be automatically controlled, recorded, stored and printed.
It has the following features: Excellent stability and convenient operation. Application: This series of products suit for various kinds of microbiology lab, pilot and industrial scale fermentation. It is ideal equipment for two step pilot scale fermentation. User can choose appreciate model according to client detail requirement.
Model | RTY-MZ | ||||||
Working volume | 5-30000L | ||||||
Filling volume | 65-80% | ||||||
Material | Wetted parts are SS316L, and others is SS304. | ||||||
Surface finish |
inside surface is 0.28-0.6μm by mechanical polishing, ( passivation or electro-polish are optional ) outside surface is 0.4-0.8μm by satin finish. |
||||||
Pressure |
Working pressure inside vessel:0.2mpa Design pressure inside vessel: 0.25mpa Working pressure inside jacket:0.3mpa Design pressure inside jacket: 0.35mpa Pressure can be also designed according to client's detail requirement. |
||||||
Agitation |
1.Fermentor use SEW decelerator 2.top mechanical agitator with two layer six blade impeller, one layer defoamer and four baffles. 3.mechanical seal (single or double mechanical seal) 4.speed range: 50-400rpm (it depends on the size of the vessel) |
||||||
Equips of vessel |
1.using alcohol flame inoculation 2.sterile sampling vale and bottom valve 3.PH, DO, temp. sensor on lower side, side glass, top with foam and level probe, and 12v sight glass with lamp, etc. |
||||||
Construction |
Floor-type Two fermentor vessels with one control system (simens PLC and HMI) They can work independently and can work together at same time. |
||||||
Operating panel | Chinese and English version | ||||||
Main components and controls of system | |||||||
PH |
1.On-line check and measure, and digitization fully automatic control; 2.PH, acid and base gang control. Peristaltic pump automatically adjust to feed acid and base; 3.Control range 2-12pH, measure precision±0.01 pH, control precision±0.1 pH; 4.It has auto alarm function. 5.pH sensor and signal transmission line, Mettler toledo |
||||||
DO |
1.On-line check and measure, cascade with rpm; digital display. 2.Range 0-100% or 0-200%, precision ±1%; 3.It has auto alarm function; 4.DO sensor and signal transmission line, Mettler toledo |
||||||
Foam control |
1.Automatically control and automatically alarm; 2. Stainless steel electrode check, peristaltic pump automatically add defoamer. |
||||||
Feed control |
1. Automatically control and automatically alarm; 2.Stainless steel electrode check, peristaltic pump automatically feeding |
||||||
Temp. control |
1.PID fully digitization signal processing, digitization control settings 2.Circulating water automatically control temperature 3.Measure range 0-150, control range 15-45,control precision±0.2 4.Pt 100Ωtemp. Sensor 5.system with electrical heater to control the temp inside vessel 6.water pump circulation 7.solenoid valve to control |
||||||
Rotate speed |
1. Automatically control , with auto alarm function 2. Siemens frequency converter to adjust the speed. |
||||||
Pressure |
1 Manually adjusting with pressure gauge and manual valve (standard system) 2. Automatically adjusting with pressure transmitter and PID valve and pneumatic valves (optional) |
||||||
Air flow |
1.Manually adjusting with rotor meter and manual valve (standard system) 2.Automatically adjusting with mass-flow meter and pneumatic valves (optional) 3.Air in filter , tail gas filter , with PALL cartridge, precision 0.2μm |
||||||
Sterilization |
1.it can be SIP by connecting with steam 2.Manually control with manual valve (standard one) 3.Automatic control with pneumatic valves (optional) |
||||||
Control | Siemens PLC, and Siemens 10.4" LCD touch screen(HMI) and RITAI program | ||||||
PC |
1.PC with Kingview software for recording function (optional) 2.PC with Siemens remote control and Kingview recording function (optional) 3.printer (optional) |
Dispenser Plug Seal,Rubble Seal Ring,Knife Gate Valve,O Ringing Seal
Dongguan Guo Hao Seals Technology Co;Ltd , https://www.ghrubberseals.com